1. In accordance with the requirements of the Ministry of Health's disinfection and sterilization technical specifications, the endoscopic instruments and their accessories (cleanable parts, handles, sleeves, etc.) after use are immediately soaked in a cleaning agent containing lysozyme for 2 to 3 minutes (dissolve Enzyme ratio product description). Lysozyme mainly digests protein and fat, eliminates agglomerates in the pipeline and organophosphorus tablets in the accessories. Achieve the effect of fast and efficient cleaning.
2. After the instrument is soaked, take it out and rinse it with clean water. The tube cavity is repeatedly rinsed with a high-pressure flushing gun. The tooth groove of the clamp end is used to clean the blood stains and the water is wiped with a soft cloth. The water drops in the tube cavity should be blown dry. Or blot it dry, and then apply lubricant. Or use an ultrasonic cleaner to clean
be careful:
(1) In the case of Hbsag positive and other infected patients, the instruments after surgery should be disinfected and then cleaned to avoid the spread of infection.
(2) When cleaning various types of pliers and scissors, care should be taken to clean the blood stains on the overlapping surface of the joint.
(3) During the cleaning process, care should be taken to keep the disassembled small parts to prevent loss.
(4) Do not immerse the wire interface connected to the machine in the cleaning fluid to prevent leakage during use. Such as the wiring port of the operating tip of the hysteroscope electric resection instrument.
(5) Take care to protect the mirror surface of the endoscope. The mirror surface of the fiber endoscope is wiped with cotton balls, and the mirror surface of the electronic mirror is wiped with gauze.
(6) The bipolar coagulation clamp should be installed at any time after disassembly and cleaning to prevent the clamp core from breaking. The ceramic port of the tip sleeve of the hysteroscope electric resection instrument should be prevented from collision during cleaning to prevent collision and cracking.